Power Supply | DC 7.4V 2600mAh rechargeable lithium battery, the battery is fully charged about 8.4V |
Display Mode | 4-digit max LCD display, white screen backlight |
Measurement Instruction | LED flashing indication during measurement, LCD countdown display |
LCD Size | 128mm×75mm;LCD Display area: 124mm×67mm |
Backlight | Controllable white screen backlight, suitable for use in dim place |
Measurement Method | DC Resistance | DC resistance bipolar method |
DC resistance quadrupole method |
Earth Resistance | Earth resistance bipolar method |
Earth Resistance Three-electrode method |
Earth resistance Quadrupole method |
Earth resistance single clamp Three-electrode method (selection method) |
Earth resistance single clamp quadrupole method (selection method) |
Earth resistance double clamp method |
Soil resistivity (Wenner method) |
Earth Current | Current clamp measurement |
Earth Voltage | Average value rectification |
Test Frequency | 128Hz/111Hz/105Hz/94Hz(AFC) |
Short Circuit Test Current | AC 250mA max |
Open Circuit Test Voltage | AC 40V max |
Electrode Spacing Range | 1m~100m |
Test Line | 4 PCS: red 20m, black 20m, yellow 10m, green 10m each 1 piece |
Simple Test Line | 2 PCS: red 1.6m, black 1.6m each 1 piece |
Auxiliary Earth Rod | 4 PCS:φ9mm×230mm (about 696g) |
Current Clamp | 2 PCS:blue and black plug, 1 red and black plug each |
Current Clamp Caliber | φ70mm |
Current Clamp Size | 223mm×120mm×39mm |
Measurement Time | AC current: about 2 times/second; earth voltage: about 2 times/second; earth resistance: about 7 seconds/time |
Line Voltage | Measurement below AC 600V |
Communication Method | Bluetooth module, USB interface, storage data can be uploaded to the computer through software |
Data Hold | When Data holding : “HOLD" symbol display |
Memory | 2000 groups, flashing "FULL”symbol indicates that the storage is full |
Data Access | Data access function: “READ" symbol display |
Overflow Display | Over-range overflow function: “OL" symbol display |
Interference Test | Automatically identify the interference signal, and the “NOISE” symbol indicates when the interference voltage is higher than 5V |
Auxiliary Earth Test | With auxiliary earth resistance value test function,0.00KΩ~30kΩ(100R+rC<50kΩ,100R+rP<50kΩ) |
Alarm Function | When the measured value exceed the alarm set value,“Du-Du-Du-" Alarm prompt |
Alarm Threshold Setting Range | Resistance: 0Ω~9999Ω, voltage: 0.0V~100.0V, current: 0.000A~9.9999A |
Automatic Shutdown | Automatically shut down after 15 minutes of power on |
Working Current | Backlight: 25mA Max (separate backlight power consumption); standby: 25mA Max (the backlight is turned off after the tester turning on); measurement: 150mA Max (the backlight turn off) |
Tester Size | 280mm×260mm×160mm |
Weight | Tester: About 1799g(include battery);total weight: about 6.13kg(include accessory) |
Battery Voltage | When the battery voltage drop to about DC 7.2V±0.1V, the low battery voltage symbol is displayed to remind that the battery need to be replaced. |
Protection Level | IP65(closed) |
Overload Protection | Measuring earth resistance: AC 280V/3 seconds between C(H)-E、P(S)-ES ports |
Insulation Resistance | Above 20MΩ(500V between the circuit and the housing) |
Withstand Voltage | AC 3700V/rms(Between the circuit and the housing) |
Electromagnetic characteristics | IEC61326(EMC) |
Suitable Safety Regulation | IEC61010-1(CAT Ⅲ 300V, CAT IV 150V, pollution level 2);IEC61010-031; IEC61557-1(Earth resistance);IEC61557-5(Soil resistivity);JJG 366-2004(Earth Resistance Tester);JJG 1054-2009(Clamp Earth Resistance Tester) |
Working Temperature & Humidity | -10℃~40℃; below 80%RH |
Storage Temperature & Humidity | -20℃~60℃; below 70%RH |